F3 COMPACT Operations Manual
25
NOTE
Maximum sensitivity is only available 30 seconds after the Threshold
Tone commences. Do not test the detector with the Test Piece until
30 seconds after the Threshold Tone begins. If the Test Piece cannot
be heard, conduct an Audio Reset and recheck. If the Test Piece
fails to be detected the detector must be considered faulty and must
not be used.
Figure 16: Test Piece procedure
NOTE
The Test Piece not only ensures that the sensitivity of the detector
is correct but also gives the operator an example of how a minimum
metal mine might sound when deeply buried; for example Type 72A
at 15 cm. (Sensitivity knob set to Position 4)
Vedere la pagina 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 43 44

Commenti su questo manuale

Donaldacith 18 Feb 2024 | 19:06:39

Welcome to the world of Premium Mycelium Daily Basis Gummies, your daily dose of fitness packed into a tasty and easy gumdrop format. Tapping into the potency of natural most effective toadstools, our chews are especially formulated to support your overall healthiness and vitality. Key Features: Premium Ingredients: Our gummies are made with the top grade components, including a combination of ultra toadstools known for their well-being perks, such as Reishi, Lion's Mane, and Cordyceps.